- Model : DSC-F8DA; DSC-F12DA
- Sampling Channels : 8 / 12
- Temperature Range : RT-50.0 ℃
- Maximum Sampling Times : 40 times
- Number of Dissolution Vessel Positions : 8 / 14
- Sampling Volume (Refilling Volume) : 1~20 mL
- Speed Range : 0–400±1 rpm / 400–1000±4% rpm
Fully Automatic Dissolution Sampling And Collection System
Warning: call_user_func_array() expects parameter 1 to be a valid callback, function 'wpb_pcf_contact_form_button' not found or invalid function name in /www/wwwroot/www.scitekuae.com/wp-includes/class-wp-hook.php on line 341
Feature:
- Integrated computer OS, smartphone-style icon interface for intuitive operation.
- 14-inch LCD all-in-one touchscreen, flexible extension and adjustable positioning.
- Four-level user authority management (Super Admin, Admin, Technician, Visitor) for structured access control.
- PEEK automatic sampling needle, corrosion-resistant, no adsorption, lifetime durability.
- 316L stainless steel paddle and basket, optional PTFE-coated paddle available.
- Automatic synchronous dosing device for simultaneous media addition in dissolution tests.
- 14-cup / 12-paddle dual-row system (DSC-F12DA), automatic sampling of up to 12 vessels with blank and standard compliance.
- Dual speed modes (synchronous & step), optional dual motors for zonal speed control.
- 8-cup / 8-paddle dual-row system (DSC-F8DA), automatic simultaneous sampling of up to 8 vessels, supporting blank and standard controls.
Specification:
| Model | DSC-F8DA | DSC-F12DA |
| Dissolution System | ||
| Number of Dissolution Vessel Positions | 8 | 14 |
| Speed Range | 0–400 ± 1 rpm / 400–1000 ± 4% rpm | |
| Temperature Range | RT-50.0 ℃ | |
| Temperature Stability | ≤±0.1 ℃ | |
| Paddle Radial Run-Out | ≤0.5 mm | |
| Basket Radial Run-Out | ≤0.8 mm | |
| Deviation Between Rotating Rod and Dissolution Cup Axis | ≤1.0 mm | |
| Power Supply | AC 220 V±10%, 50 Hz, 1800 W | |
| Sampling System | ||
| Sampling Channels | 8 | 12 |
| Maximum Sampling Times | 40 times | |
| Sampling Volume (Refilling Volume) | 1–20 mL | |
| Minimum Sampling Interval | 1 minute for the first time, 2 to 5 minutes for the rest
(depending on the sampling volume and whether fluid is replenished) |
|
| Maximum Working Time | 1 to 100,000 minutes (69 days) | |
| Sampling Accuracy | ≤±1 % (average value of each channel for 5 mL sampling volume) | |
| Sampling Filtration Cycle | <30 s (fully in line with the requirements of the Chinese Pharmacopoeia) | |
| Dimensions (W×D×H)(mm) | 550×900×810 | 600×1400×830 |
| 620×550×536 | 620×550×536 | |
| N.W. (kg) | 80/60 | 150/60 |