Application:
The microbial air sampler is designed and manufactured in accordance with the requirements of the latest version of China’s GMP regulations, based on the principle of the fifth level of Andersen impact classification, and with reference to domestic similar products. The shell is made of anodized aluminum which is corrosion-resistant and easy to wipe—features that facilitate cleaning and sterilization. The sampling head of the microbial air sampler is made of anodized aluminum material, with a total of 397 holes, 0.6mm – diameter micropores (straight holes, without chamfering), reducing bacterial overlap and lowering the error in microbial counting.
Advantage:
- Intuitive Operation: The color-screen design makes the appearance simple and elegant, and the operation interface is clear and straightforward.
- Memory Function: It can preset 100 sets of points, enabling quick sampling as needed.
- Reminder Function: It has a reminder function for historical data deletion to prevent incorrect operations.
- Time-setting: It comes with a standard-equipped interval time setting and delay time setting to improve the accuracy of detection results.
- Real-time Data Printing: Optional Bluetooth printer can be equipped to print data in real-time, facilitating the archiving of detection data.
- Compliance with Standards: It meets the requirements of the latest version of GMP and can also satisfy those of most established industry standards.
- High-quality Core Components: It adopts an imported air pump, featuring high precision and a long service life.
- Meet Detection Requirements: It can be equipped with a high-pressure separator to meet the requirements for monitoring compressed air in clean rooms.
Specification:
| Model |
AMS-IV |
| Screen |
4.3-inch color touch screen, operable with gloves |
| Number of Preset Sampling Points |
Can preset 100 groups of points |
| Power Supply |
AC 220 V converted to DC16.8 V |
| Sampling Flow Rate |
100 L/min ± 5% (error) |
| Sampling Delay Time |
0 – 99 s (adjustable) |
| Sampling Interval Time |
0 – 99 s (adjustable) |
| Sampling Velocity |
0.38 m/s (isokinetic sampling) |
| Impact Rate |
10.8 m/s (based on the principle of the fifth level of Andersen impact classification) |
| Sampling Volume |
1 – 9999 L (adjustable) |
| Material of Machine Body and Sampling Head |
Anodized aluminum |
| Data Storage Capacity |
10,000 records |
| Continuous Working Time |
8 h |
| External Size(W×D×H) |
147×115×190 mm |
| Package Size (W×D×H) |
363×350×180 mm |
| N.W/G.W (kg) |
1.8 / 2.5 |